Optimum Temperature for Nitrogen Evaporation Instruments

January 14, 2014 /

Product Information

/ Organomation

Nitrogen evaporation is a widely used technique for concentrating samples prior to analysis. The temperature at which the evaporation is conducted is crucial for achieving optimal results. This post discusses the importance of temperature control and provides guidelines for selecting the right temperature for your nitrogen evaporation instrument.

Temperature control is essential for efficient evaporation. If the temperature is too low, the evaporation process will be slow, leading to longer processing times. Conversely, if the temperature is too high, there is a risk of degrading the sample or losing volatile compounds. Proper temperature control ensures that samples are processed quickly and accurately, maintaining their integrity.

 

Key Factors Affecting Optimum Temperature in Nitrogen Blowdown

 

Sample Type

Different samples have varying boiling points. Selecting the appropriate temperature is essential to avoid degradation and ensure the efficient evaporation of the solvent. For example, organic solvents generally evaporate at lower temperatures compared to aqueous solutions.

 

Solvent Properties

The boiling point of the solvent being evaporated is a critical factor. Solvents with higher boiling points require higher temperatures for efficient evaporation, while those with lower boiling points need lower temperatures to prevent overheating and potential loss of analytes.

 

Instrument Design

The design and capabilities of the nitrogen evaporation instrument play a significant role in determining the optimal temperature. Advanced instruments with precise temperature control systems can handle a wider range of temperatures, providing flexibility for different applications. Instruments with poor temperature control may require more conservative settings to avoid sample damage.

 

Recommended Temperature Ranges

 

General Guidelines

For most applications, a temperature range of 30°C to 60°C is recommended. This range is typically sufficient to evaporate common solvents without causing thermal degradation of the samples. However, the specific requirements may vary based on the sample and solvent properties.

 

Organic Solvents

For organic solvents, temperatures between 40°C and 50°C are generally effective. This range allows for efficient evaporation while minimizing the risk of degrading heat-sensitive analytes.

 

Aqueous Solutions

Aqueous solutions often require slightly higher temperatures, ranging from 50°C to 60°C. This helps to overcome the higher boiling points of water and other components in the solution.

 

Heat-Sensitive Samples

For heat-sensitive samples, it is advisable to use temperatures below 40°C. Lower temperatures help prevent any damage to the sample components, ensuring the integrity of the analytes is maintained.

 

Detailed Considerations for Nitrogen Evaporation Instruments

In a nitrogen evaporation instrument, temperature and gas flow rate are the two major operational parameters that determine the rate at which the solvent is removed from a system. Adjusting the bath temperature can have varying effects on the analyte, and it is important to remember that a higher temperature is not always optimal. In many cases, it may be necessary to sacrifice time in favor of sample integrity. The investigator must determine the optimal combination of bath temperature and gas flow rate that will remove the unwanted solvent without degrading the sample.

 

The Role of Temperature

Temperature is the operational parameter most often used to control the evaporation rate of the solvent from the sample. An increase in the bath temperature increases the evaporation rate of the solvent. An operational bath temperature approximately 10°C below the boiling point of the solvent being removed is often considered optimal. In some cases, however, the analyte or substance of interest may be susceptible to thermal degradation at or below the boiling point of the solvent in which it is contained, thus limiting the operational temperature range. Reducing the operational bath temperature to lower the possibility of thermal degradation is a sound approach but must be weighed against the possibility that longer exposure at a lower temperature may be just as detrimental as a shorter exposure time at a higher temperature.

 

Effects of Chemical Reactions

In addition to thermal degradation, there is also a possibility of chemical reactions between the sample and the other substances within the sample’s solution. The reaction rate of a system may increase or decrease with temperature depending on the reactants and reaction conditions. This means that a longer exposure time at a lower temperature could result in greater damage to the sample than a shorter exposure time at a higher temperature. The reaction rate may also be related to the concentration of one or more of the components of the system. As solvent is removed from the sample, the concentration of the sample and other nonvolatile reactants or components increases, potentially increasing the reaction rate of intra-sample reactions. Therefore, increased exposure time will always increase the effect of any degradative or intra-sample reaction processes.

 

Hypothetical Systems

 

System #1

In this system, the analyte is a polymer synthesized in toluene and quenched with methanol. The sample solution contains toluene, water, methanol, residual monomer, and a non-volatile polymer. The water bath temperature is set as close to the boiling point of toluene, the highest boiling and predominant component, as possible. The N-EVAP nitrogen evaporator water bath is set at 90°C, and the nitrogen flow at 8 LPM. Under these conditions, the solvent removal takes less than 60 minutes, and the polymer is isolated and quantified efficiently.

 

System #2

This system contains a different monomer and polymer but is processed similarly to System #1. However, no polymer is found at a higher temperature. When the solvent is removed at 50°C with a gas flow of 8 LPM, taking over 180 minutes, polymeric residue is found. The polymer is thermally unstable at the higher temperature and depolymerizes, reverting back to monomer.

 

System #3

In this system, the monomer has a boiling point of 95°C and a vapor pressure of 42 mmHg at 25°C. The solvent is removed at 50°C and 95°C, respectively, with a gas flow of 8 LPM. At 95°C, the polymer is isolated in about 60 minutes without thermal degradation, but the weight is lower than expected. At 50°C, the weight of isolated residue is above the theoretical value due to the slower evaporation rate of the unreacted monomer.

 

Conclusion

Optimizing the temperature settings for nitrogen evaporation instruments is essential for achieving accurate and reliable results. By considering factors such as sample type, solvent properties, and instrument design, users can ensure efficient evaporation while maintaining the integrity of their samples. The optimal operational parameters for safe and efficient isolation of the analyte vary for each system. It is important to note the properties of each part of the sample solution and consider the trade-offs between temperature, exposure time, and sample integrity. The effect of gas flow rate on evaporation time will be discussed in an upcoming article.