Common User Complaints of SpeedVac Style Centrifugal Evaporators

 

Users of laboratory centrifugal evaporators, particularly benchtop SpeedVac-style concentrators, report several recurring problems on online forums. These problems fall into four main categories: unintended sample freezing, sudden vacuum loss, failure to evaporate solvents, and difficulty achieving or interpreting deep vacuum.

 

Unintended Sample Freezing

Several users have noted that evaporative cooling in their SpeedVac units has actually frozen their samples in the process of concentrating them. In one LabWrench discussion, a user described how their Savant AS160-120 both evaporated and froze samples, with other users suggesting to “increase the chamber temperature at the beginning of the evaporation process”, just to prevent ice formation [2].

 

Sudden Vacuum Loss and Suction Failures 

Another frustration arises from the vacuum system failing abruptly, halting concentration. In another LabWrench thread, one researcher reported that with their second-hand SpeedVac “a vacuum isn’t forming to allow us to collect evaporated hexane,” indicating that well-used units may lose suction without warning, complicating lab processes [3].

 

Apparent Lack of Evaporation

Users sometimes find that their machine appears to function—rotor spinning, vacuum engaged, heat applied—but no solvent evaporation occurs. One SpeedVac SVC100H owner wrote that “when I turn the vacuum on and spin the rotor…there is no evaporation of 1 mL water after 2 hours”. It was also noted by this user, that even with the heat enabled, there was a lack of water evaporation, suggesting possible vacuum leaks or pump issues that may be preventing solvent removal [4].

 

Difficulty Achieving and Interpreting Deep Vacuum

A Reddit thread in r/labrats highlights a common confusion over vacuum readings and the inability to reach pressures low enough to maintain frozen samples. One user struggled to keep a pre-frozen sample solid, discovering that their SpeedVac’s vacuum gauge units were being misunderstood (e.g., mistaking mtorr for torr) [5]. They also found that “the vacuum is not pulling a high enough vacuum,” most often due to leaks or seal failures.

These collective user experiences emphasize that, despite their convenience, some centrifugal evaporators can pose significant operational challenges. This is particularly true regarding vacuum integrity and temperature contrast, which often require careful troubleshooting, routine maintenance, and sometimes hardware upgrades to resolve.

 

Nitrogen Blowdown: A Faster Alternative

The comparative data leaves little doubt: for laboratories that need to drive solvents to dryness quickly and reproducibly, nitrogen blow-down is the speed champion. Even under deliberately conservative conditions (bath temperature set at 35 °C instead of the recommended 63 °C), Organomation’s N-EVAP dried 1.5 mL of methanol 33% faster in 2 mL tubes and 38% faster in 15 mL Falcon tubes than the Thermo Scientific SpeedVac [6]. When the N-EVAP is run at its optimal 63 °C bath temperature, prior tests have shown that its evaporation rate climbs to ~0.25 mL min, roughly four times the rate observed in this head-to-head study and more than quadruple the SpeedVac’s performance [6].

Besides accelerated speed, another advantage of nitrogen blow-down is that it delivers extra throughput without the use of vacuum pumps, cold traps, rotors, or seals. These components make centrifugal systems more expensive to purchase and add to their service and repair costs. Targeted nitrogen gas needles minimize cross-contamination risks and the open-tube format lets operators monitor the endpoint visually instead of relying on vacuum gauges or timers.

In short, if your primary bottleneck is solvent removal, especially when working with volatile, heat-sensitive, or high-volume samples, nitrogen evaporation offers a faster, simpler, and ultimately more economical solution. The Organomation study underscores what many high-throughput labs have already discovered: switching from centrifugal evaporators to nitrogen blow-down can trim minutes (or even hours) from each run, sharpen turnaround times, and free up valuable instrument capacity for what matters most, downstream analysis [6].

 

Easily estimate how long your samples would evaporate in a nitrogen evaporator with our calculator:

 

 

 

Citations:

  1. https://blog.organomation.com/blog/types-of-laboratory-evaporators
  2. https://www.labwrench.com/thread/12064/speed-vac-is-freezing-my-samples
  3. https://www.labwrench.com/thread/248754/speed-vap-suction-failure
  4. https://www.labwrench.com/thread/174584/appears-to-work-but-no-evaporation
  5. https://www.reddit.com/r/labrats/comments/1bc2mb6/comment/kujq20v/https://www.reddit.com/r/labrats/comments/1bc2mb6/comment/kujq20v/
  6. https://blog.organomation.com/blog/evaporating-methanol-to-dryness-is-centrifugal-or-nitrogen-blowdown-faster